Bimetal Thermometer with Thermocouple
WSS-481NH2 bimetallic thermometer with thermocouple (resistance) and temperature transmitter is a temperature sensor for local display and remote signal transmission. It can indicate local temperature as well as transmit two-wire system signal of thermoco
Technical parameters
Measuring range: -60 ~ 500 ℃
Display accuracy: 1.5
Remote signal deviation:
A. Graduation: E (nickel chromium-copper nickel) allowable deviation ± 2.5 ℃ or 0.75% t
B. Graduation: K (nickel chrome-nickel silicon) allowable deviation ± 2.5 ℃ or 0.75% t
C. Graduation: Pt100 platinum resistor () allowable deviation:
Class A: ± (0.15 + 0.002 | t |)
Class B: - (0.30 + 0.005 | t |)
D. Basic error with temperature transmitter: Δ = Δ 1 + / - 0.5% FS
Δ 1 is thermocouple (resistance) basic error, FS is measuring range
Temperature transmission: two-wire system (4 ~ 20 mA), transmitter minimum working voltage is 12V, maximum 35 V, rated voltage 24V
Work environment:
A. environmental temperature: -25 ℃ ~ 80 ℃ (not higher than 70 ℃ in dangerous place)
B. relative humidity: 5% ~ 95%
C. mechanical vibration: f≤55Hz, amplitude < 0.15 mm
D. no corrosive medium in the environment
Environment temperature influence: ≤0.05%/1℃
Model
W |
S |
S |
Specifications |
|||||||
Temperature Instrument |
Metal expansion type |
bimetallic |
- |
|||||||
4 |
Φ100mm |
|||||||||
5 |
Φ150mm |
|||||||||
Dial nominal diameter |
0 |
Angle or axial type |
||||||||
1 |
Straight or radial type |
|||||||||
8 |
Universal type |
|||||||||
structural style |
1 |
Movable male thread |
||||||||
2 |
Movable female thread |
|||||||||
3 |
Movable Flange |
|||||||||
Mounting type |
thermocouple |
|||||||||
E |
nickel chrome- Copper nickel, type E |
|||||||||
N |
Nickel chrome- nickel silicon, type K |
|||||||||
Thermal resistance |
||||||||||
Pt100 |
Platinum resistance |
|||||||||
Cu50 |
Copper Resistance |
|||||||||
Sensor |
Without temperature transmitter |
|||||||||
H2 |
The output and the temperature is in linear |
|||||||||
H4 |
Output and input signal is in linear |
|||||||||
temperature transmitter |
||||||||||
W |
S |
S |
- |